Product name | industrial resistance meter | |||||
---|---|---|---|---|---|---|
Model | HE-480R | |||||
Measurement method | Electrode type (2-electrode method) | |||||
Cell constant | 0.01/cm | |||||
Temperature sensor specifications | Platinum resistance 1000 Ω/0℃ | |||||
Measuring range | Measurement unit | kΩ•m | 0 to 2.00 | 0 to 20.0 | 0 to 200.0 | 0 to 1000 * |
MΩ•cm | 0 to 0.200 | 0 to 2.00 | 0 to 20.00 | 0 to 100.0 * | ||
*: Measurable without temperature compensation | ||||||
Temperature: 0℃ to 100℃ (Select your desired decimal point from 0, 1, and 2 digits.) | ||||||
Repeatability | Within ±0.5% of the full scale (in equivalent input) | |||||
Linearity | Within ±0.5% of the full scale (in equivalent input) | |||||
Transmission output | 4 mA to 20 mA DC: input/output isolated type Maximum load resistance: 900Ω Transmission output range: Freely selectable within the measurement range | |||||
Contact output | Outputs: 2 points Alarm contact output (R1 and R2) Contact type: relay contact, SPDT (1c) Contact rating: 240 V AC, 3 A and 30 V DC, 3 A (resistance load) Contact function: selectable from upper/lower limit operation (ON/OFF control), alarm, and maintenance. | |||||
Calibration function | Specific resistance: Based on the specified compensation coefficient for cell constant (parameter input) Temperature: Calibrated by comparing with the reference thermometer | |||||
Transmission output hold feature |
| |||||
Self-diagnosis function |
| |||||
Temperature compensation |
| |||||
Temperature compensated range | 0℃ to 100℃ | |||||
Extra deionized water specific resistance selection | Measurement unit | kΩ•m | 182.3 (standard), 181.8, 182.4 | Select from options shown on the left. | ||
MΩ•cm | 18.23 (standard), 18.18, 18.24 | |||||
Clipping function | When the measured value is above the upper limit of the measurement range derived from the specified specific resistance, the specified resistance is used as the measured value. | |||||
Ambient temperature | −5℃ to 45℃ | |||||
Relative humidity | 20% to 85% (without dew condensation) | |||||
Storage temperature | −25℃ to 65℃ | |||||
Power supply | Rated voltage 100 V to 240 V AC, 50/60 Hz, 10 VA (max.) | |||||
Structure | Indoor-use panel installation type Panel case: ABS, Terminal: PBT Panel: IP65 dust and water proof structure | |||||
Protective structure | Panel: IP65 (IEC60529, JIS C0920) Rear case: IP20, Terminal:IP00 Class II device (IEC61010-1) Pollution level 2 (IEC61010-1) | |||||
Conforming standards | CE Marking | EMC Directive: EN61326-1 Low Voltage Directive: EN61010-1 RoHS Directive: EN50581 | ||||
FCC Rule | FCC Part15 | |||||
External dimensions | 48(W) mm × 96 (H) mm × 115 (D) mm Case depth: approx. 105 mm (when panel-mounted) | |||||
Mass | Approx. 400 g | |||||
Compatible sensors | ERD/ERF-series specific resistance sensor, cell constant 1.0/m (0.01/cm) |