Principle of measurement
CLD | NDIR | NDUV | MPA | |
---|---|---|---|---|
NOX | ● | ● | ||
SO2 | ● | ● | ||
CO | ● | |||
CO2 | ● | |||
O2 | ● | |||
NH3 | ● |
CLD: Chemiluminescence detection
NDIR: Non-dispersive infrared
NDUV: Non-dispersive ultraviolet
MPA: Magneto-pneumatic
Ranges
Component | Principle | Range | |
NOX (Low) | CLD | 0-5/5000 ppm (dual range) | |
NOX | NDIR | 0-200/5000 ppm (dual range) | |
SO2 (Low) | NDUV | 0-10/1000 or 0-5/1000 ppm | |
SO2 | NDIR | 0-200/5000 ppm (dual range) | |
CO (Low) | NDIR | 0-5/500 ppm (dual range) | |
CO | NDIR | 0-100/5000 ppm (dual range) | |
CO (High) | NDIR | 0-5 vol% | |
CO2 | NDIR | 0-5/50 vol% | |
O2 | MPA | 0-5/25 vol% | |
NH3 | CLD | 0-10/100 ppm (dual range) |
Repeatability | ±0.5% full-scale |
Zero drift | ±1.0% full-scale per week |
Span drift | ±2.0% full-scale per week |
Linearity | ±1.0% full-scale per week |
Communications | RS-232C, RS-485 |
Serial Ports | USB, Ethernet |
Ambient temperature | Standard: +5°C to 40°C, |
Response time | 45 seconds max. at system inlet (T90) |
Total interference | ±2.0% full-scale |
Dimensions | [inches] 25.6 (W) x 32 (D) x 75 (H) |
Weight | Approx. 700 lb (318 kg) |