General Compact Raman Spectrometer Specifications | |
---|---|
Spectral Coverage | 785 - 1060 nm (Also available: 532 nm optimized spectrometer) 150 to > 3300 cm-1 (Note: low limit depends on Raman probe) |
Spectral Resolution (at F/6) | < 0.6 nm from 780 nm to 965 nm < 1.5 nm from 965 nm to 1,060 nm 6 cm-1 at 1777 cm-1 7 cm-1 at 2518 cm-1 |
Focal Length | 140 mm |
Optical Dispersion | 9.77 nm/mm; 0.137 nm/pixel |
Stray Light | < 0.1% (typical) |
Wavelength Accuracy | 0.025 nm |
F# | F/2.3 (for highest sensitivity) F/6 (for the most compact and highest resolution configuration) |
Software | Basic LabVIEW acquisition software for evaluation. See our optional LabSpec software. (DLLs provided for software integration) |
OEM SYNCERITY™ Scientific Camera with 2048 x 70 – VIS-NIR Optimized CCD* | |
---|---|
Quantum Efficiency at 20° C | New VIS-NIR Optimized CCD (Also available: 532 nm optimized spectrometer) |
CCD Sensor Format | 2048 x 70 |
CCD Pixel Size | 14 x 14 µm |
CCD Height | 1 mm |
CCD QE at 20° C | 60% at 500 nm; 80% at 600 nm; 80% at 800 nm 68% at 900 nm; 42% at 1000 nm; 20% at 1,060 nm See curve below |
Deep Thermoelectric Cooling | -50° C @ +25° C ambient or -40° C @ +40° C ambient. Yields low dark current suitable for most OEM and many research applications |
Single Pixel Well Capacity | 50,000 e¯/pixel (minimum); 60 000 e¯/pixel (typical) |
Serial Register Full Well Capacity | 250,000 e¯/pixel (minimum); 500 000 e¯/pixel (typical output register saturation) |
Scan Rates | 45 kHz and 500 kHz |
Readout Noise (at 45 kHz and at –45° C) | 10 e¯ (typical) to 12 e¯ (maximum) |
Readout Noise (at 500 kHz and at –45° C) | 20 e¯ (typical) to 25 e¯ (maximum) |
Dynamic Range (typical) | 55,500:1 |
Maximum Spectral Rate | 20 Hz at 45 kHz scan rate; 189 Hz at 500 kHz scan rate |
Digitization | 16-bit ADC |
Non-linearity (measured on each camera) | <0.15% (typical) at 45 kHz (0.4% maximum) <0.20% (typical) at 500 kHz (1% maximum) |
Dark Current at –45° C Note: Pixel Size = 14 μm | 0.025 e–/pixel/s (typical) |
Software-adjustable Gains | 2, 4, and 10 e–/count at –50° C |
Communication | USB 2 |
Environmental Conditions | Operating temperature 0° C to 40° C ambient Relative humidity <70% (non-condensing) Storage temperature -25° C to 45° C |
Power Requirements AC/DC Power Supply (provided) | 90-264 VAC, 47–63 Hz |
Quantum Efficiency of the CCD Detector
Mechanical Drawings
You can email OEM Technical Sales at: OEM.US(at)horiba.com or contact OEM at: +1.732.494.8660 Ext. 7733