InVizU Cube

InVizU Cube
InVizU Cube optical layout
InVizU Cube spectral coverage
InVizU Cube Quantum Efficiencies

Ultra Compact and Broad Range Spectrometer

InVizU Cube (Infrared-Visible-Ultraviolet) is the latest evolution of the ultra compact, high performance spectrometers from the HORIBA Scientific OEM team. This dual system for industrial applications uses a modified miniaturized VS70 optical engine optimized for the UV-VIS spectral range along with the VS20 optimized for the NIR spectral range. The InVizU Cube is based on high performance aberration corrected concave holographic gratings fitted with custom multi-area order-sorting filter to eliminate higher orders.

This platform is specially designed to easily adapt to a large variety of detectors and electronic drivers. The optical design is optimized to minimize stray light and maximize optical performance. The InVizU Cube outperforms current generation of broad range mini-spectrometers thanks to its compact sleek design, high sensitivity, flat spectral response and resolution.

For OEM Industrial Applications

  • Fluorescence
  • Photoluminescence
  • Reflectance
  • Emission
  • Examples:
    • Semiconductors
    • Food Integrity and Pharma Production
    • Life Sciences
    • Analytical Chemistry
    • Metrology
    • Process Monitoring

 

*Available in select countries.

You can email OEM Technical Sales at: OEM.US(at)horiba.com or contact OEM at: +1.732.494.8660 Ext. 7733

Segment: Scientific
Manufacturing Company: HORIBA Scientific
  • Extended spectral coverage of 200 -1700 nm
  • Super compact size (100 x 100 x 108 mm) robustness and stability 
  • High SNR CMOS sensor combined with cooled NIR IGA array 
  • Simultaneous acquisition of  UV-VIS and NIR spectra
  • Includes a custom-madebifurcated SMA fiber as a dual input to the spectrometer
  • Triple USB 2.0 Ethernet and EtherCAT communication
  • High throughput and ultra low stray light

 

 

OEM Spectrometer Selection Guide

View and Download our OEM Spectrometer Selection Guide

You can email OEM Technical Sales at: OEM.US(at)horiba.com or contact OEM at: +1.732.494.8660 Ext. 7733

 

General Spectrometer Specifications*

Spectral Coverage

200 - 1,700 nm

Spectrometer Input

Input port: SMA for UV-VIS and custom-end ferrule for NIR (see drawing for details)

Spectral Resolution

< 6 nm
Average Spectral Dispersion

17 nm/mm for UV-VIS and 75 nm/mm for NIR

Focal Length

70 mm for UV-VIS and 20 mm for NIR

F/#

F/4 for UV-VIS and F/2.7 for NIR

Stray Light

± 0.1 nm

Wavelength Accuracy

< 0.4 nm (using multi-area wavelength calibration) for UV-VIS and 5 nm for NIR

Software

LabVIEW™ acquisition software for initial evaluation (DLLs provided for software integration)

Detector Options and Specifications*
 High Speed Linear Front Illuminated CMOS SensorTE Cooled Linear InGaAs Detector
Detector ModelHamamatsu CMOS lineari mage Sensor S11639 with active pixel structure and global e-shutterHamamatsu InGaAs Linear Image Sensor G11508
on a CMOS chip
Detector Sensor Format2048 x 1 pixels512 x 1 pixels (256 x 1 pixels, optional)
Detector Active Area28.7 mm x 0.2 mm12.8 mm x 0.5 mm
Detector Pixel Size14 μm x 200 μm25 μm x 500 μm (60 μm x 500 μm, optional)
Detector QERefer to graphs belowRefer to graphs below
Full Well Capacity100,000 e¯ (typical)
80,000 e¯ (minimum)
Low-Gain: 175,000,000 e¯
(typical)
High-Gain: 17,500,000 e¯
(typical)
Readout Noise16 e¯ (typical); 20 e¯ (maximum)Low-Gain: 16,000 e¯
(typical)
High-Gain: 1,900 e¯ (typical)
Dynamic Range
(FW/RN)
6,250:1 (typical)
75.9 dB (typical)
Low-Gain:11,000:1 (typical)
80.8 dB (typical)
High-Gain: 9,000:1 (typical)
79.1 dB (typical)
SNR (SQRT FW)316Low-Gain: 13,228High-Gain: 4,183
Dark Current375e¯ pixel/s (typical)
500 e¯/pixel/s (maximum) @ 20° C
Within ±250,000 e¯/pixel/s (typical) @ 5° C
Maximum Spectral Rate5,000 spectra/s (Single IGA Sensor Mode)
833 spectra/s (Single CMOS Sensor Mode, or Dual Sensor Mode of CMOS+IGA)
ADC16-bit
Thermoelectric Cooling
(1-Stage)
Uncooled0° C @ T (ambient external) ≤ +30° C
+5° C @ T (ambient external) ≤ +35° C
+10° C @ T (ambient external) up to +40° C
Non-linearity<1% (corrected)<1% (uncorrected, will be smaller after correction)
Sensor Operating ModesSingle CMOS Sensor (UV – VIS)
Single IGA Sensor (NIR)
Dual Sensor (CMOS & IGA)
Exposure Time20 μs up to 100 sec (defined only by dark current)
Exposure Time Modes
for Dual Sensor Operation
Standard (equal sensor integration time values) and variable
(different synchronized sensor integration time values)
Communication InterfaceMulti Communication (MC): USB 2.0 HS, Ethernet (10/100), or EtherCAT (Optional)
Environmental ConditionsOperating temperature 15° C to 40° C ambient
Relative humidity <70% (non-condensing)
Storage temperature -25° C to 45° C
Power Requirements
AC/DC Power Supply
(provided)
24VDC @ 0.36A typical (3-Pin M8 Connector)


* HORIBA Instruments has a policy of continuous product development, and reserves the right to amend part numbers, descriptions and
specifications without prior notice

 

Custom Fiber Drawing

Drawing of a custom one-to-dual input fiber designed to maximize light collection into the InVizU Cube two detector spectrometer

Quantum Efficiencies

 

 

Mechanical Drawings

Download Mechanical Drawings

 

You can email OEM Technical Sales at: OEM.US(at)horiba.com or contact OEM at: +1.732.494.8660 Ext. 7733

 

Request for Information

Do you have any questions or requests? Use this form to contact our specialists.

* These fields are mandatory.