General Spectrometer Specifications* | |
---|---|
Spectral Coverage | UV-VIS (200–800 nm) with 477 l/mm grating, VIS (280--700 nm) with 582 l/mm grating, UV-NIR (200–985 nm) with double blazed 365 l/mm grating. On request, other spectral ranges available; different mechanical integration might be needed |
Spectrometer Input | Fiber-coupled: FC or SMA; Free space: 12-25-37-50-62-75-100-125-150-200 μm slits; Other options available upon request |
Spectral Resolution | 2 nm for 365 l/mm grating, 1 nm for 477 l/mm grating, 0.8 nm for 582 l/mm grating (25 μm slit for all) |
Average Spectral Dispersion | 30.9 nm/mm for 365 l/mm grating, 23.7 nm/mm for 477 l/mm grating 16.3 nm/mm for 582 l/mm grating |
Average Digital Resolution with 25 (50) μm pixel width | 0.8 (1.5) nm/px for 365 l/mm grating, 0.59 (1.2) nm/px for 477 l/mm grating, 0.4 (0.8) nm/px for 582 l/mm grating |
Focal Length | 70 mm |
Options | Selection of high grade sensors: Hamamatsu® S1590x and S1012x families (other PDA sensors available upon request); Input port: SMA, FC, free space, custom input (e.g., round-to-line fiber bundle) |
F/# | F/2.8 |
Stray Light Rejection Typical | 0.01% for UV-VIS configuration measured at 400 nm and 700 nm with a 550 nm (40 nm band) bandpass filter, 31 μm slit-width |
Wavelength Accuracy | 0.2 nm |
Software** | LabVIEW™ acquisition software for initial evaluation (DLLs provided for software integration) |
Detector Options and Specifications*
OPTION A (High SNR, High Spectral Resolution) | |
---|---|
Detector Model | Hamamatsu S10124-1024 o r S15909-1024 with high UV sensitivity. |
Sensor Format | 1024 x 1 pixels |
PDA Pixel Size | 25 μm x 2500 μm |
PDA Active Area | 25.6 mm x 2.5 mm |
PDA Spectral Response | S15909-1024 has improved spectral response (NO FRINGES) compared to S10124-1024. (See figure below) |
Sensor Temperature | Uncooled |
Full Well Capacity | 624 Me- for S15909-1024 and 468 Me- for S10124-1024 |
Readout Noise | 14,700 e- (typical) for S15909-1024. S10124-1024 not yet evaluated |
Digitization | 18-bit |
Dynamic Range (FW/RN) | 42,500:1 (typical) for S15909-1024. S10124-1024 not yet evaluated |
Max SNR (√FW) | 24,980:1 (typical) for S15909-1024 and 21,600:1 (typical) for S10124-1024 |
Non-linearity (measured on each system) | < 0.4% (corrected) |
Dark Current (at 25C ambient) | 190,000 e-/pix/s (typical) for S15909-1024 and 625,000 e-/pixel/s (typical) for S10124-1024 |
Communication | USB 2.0 High-speed, (USB 3.0 customizable) |
Environmental Conditions | Operating temperature 15° C to 40° C ambient Relative humidity <70% (non-condensing) Storage temperature -25° C to 45° C |
Power Requirements | 5 VDC @ 0.13 A, USB-powered or Molex connector |
OPTION B (High SNR, Medium Spectral Resolution) | |
---|---|
Detector Model | Hamamatsu S15908-512 or S10121-512 with high UV sensitivity. |
Sensor Format | 512 x 1 pixels |
PDA Pixel Size | 50 μm x 2500 μm |
PDA Active Area | 25.6 mm x 2.5 mm |
PDA QE | S15908-512 has improved spectral response (NO FRINGES) compared to S10121-512. (See figure below) |
Sensor Temperature | Uncooled |
Full Well Capacity | 1.25 Ge- for S15908-512 and 1.03 Ge- for S10121-512 |
Readout Noise | 14,700 e- (typical) |
Digitization | 18-bit |
Dynamic Range (FW/RN) | 42,500:1 (typical) |
SNR | 24,000:1 (typical) |
Non-linearity (measured on each system) | < 0.4% (corrected) |
Dark Current | 190,000 e-/pix/s (typical) |
Communication | USB 2.0 High-speed, (USB 3.0 customizable) |
Environmental Conditions | Operating temperature 15° C to 40° C ambient Relative humidity <70% (non-condensing) Storage temperature -25° C to 45° C |
Power Requirements | 5 VDC @ 0.13 A, USB-powered or Molex connector |
OPTION C (Medium SNR, High Spectral Resolution) | |
---|---|
Detector Model | Hamamatsu S10123-1024 with high UV sensitivity |
Sensor Format | 1024 x 1 pixels |
PDA Pixel Size | 25 μm x 500 μm |
PDA Active Area | 25.6 mm x 0.5 mm |
PDA QE | (See figure below) |
Sensor Temperature | Uncooled |
Full Well Capacity | 87 Me- |
Readout Noise | 14,700 e- (typical) |
Digitization | 18-bit |
Dynamic Range (FW/RN) | Not yet evaluated. |
SNR | 9,330:1 (typical) |
Non-linearity (measured on each system) | < 0.4% (corrected) |
Dark Current | 125,000 e-/pix/s (typical) |
Communication | USB 2.0 High-speed, (USB 3.0 customizable) |
Environmental Conditions | Operating temperature 15° C to 40° C ambient Relative humidity <70% (non-condensing) Storage temperature -25° C to 45° C |
Power Requirements | 5 VDC @ 0.13 A, USB-powered or Molex connector |
OPTION D (Medium SNR, Medium Spectral Resolution) | |
---|---|
Detector Model | Hamamatsu S10121-512 with high UV sensitivity. |
Sensor Format | 512 x 1 pixels |
PDA Pixel Size | 50 μm x 500 μm |
PDA Active Area | 25.6 mm x 0.5 mm |
PDA QE | (See figure below) |
Sensor Temperature | Uncooled |
Full Well Capacity | 200 Me- |
Readout Noise | 14,700 e- (typical) |
Digitization | 18-bit |
Dynamic Range (FW/RN) | Not yet evaluated. |
SNR | Not yet evaluated. |
Non-linearity (measured on each system) | < 0.4% (corrected) |
Dark Current | 125,000 e-/pix/s (typical) |
Communication | USB 2.0 High-speed, (USB 3.0 customizable) |
Environmental Conditions | Operating temperature 15° C to 40° C ambient Relative humidity <70% (non-condensing) Storage temperature -25° C to 45° C |
Power Requirements | 5 VDC @ 0.13 A, USB-powered or Molex connector |
* HORIBA Instruments has a policy of continuous product development, and reserves the right to amend part numbers, descriptions and
specifications without prior notice.
**No LabVIEW™ license is needed to run our acquisition software.
PDAs Spectral Response
Mechanical Drawings
You can email OEM Technical Sales at: OEM.US(at)horiba.com or contact OEM at: +1.732.494.8660 Ext. 7733